Dynamic compression plate

A dynamic compression plate (DCP) is a metallic plate used in orthopedics for internal fixation of bone,[1] typically after fractures. As the name implies, it is designed to exert dynamic pressure between the bone fragments to be transfixed. Dynamic compression is achieved either by attaching a tension device to a plate or by using a special dynamic compression plate.[2] However, compression plating with a tension device requires a longer surgical incision to allow insertion of the device, and plating carries the possibility of refracture after the plate is removed.[3] A neutralization plate is used to bridge a comminuted fracture; it also transmits bending or torsional forces from the proximal to the distal fragment. Plates used for buttressing prevent collapse by supporting an area of thin cortex or cancellous bone graft.[4]
When plates are used, atrophy (osteopenia) of the bone beneath the plate may predispose the bone to fracture.[5] After removal of plates, fractures may occur through the screw holes.[6][7]
Sliding screw plate
The sliding screw plate (dynamic compression screw, dynamic hip screw) may be used to treat intertrochanteric fractures as well as other injuries. This device consists of a lag screw and a side plate with a barrel. The sliding screw telescopes and provides fixation while allowing impaction to occur at the fracture during healing and weight bearing.[8] Among the complications of sliding screw plate fixation are "cutting out" of the screw, penetration of the screw into the joint,[9] bending or breaking of the implant, disengagement of the screw from the barrel,[10] and intrapelvic protrusion of the screw.[11]
See also
References
- ↑ Cohn, M. R.; Unnanuntana, A.; Pannu, T. J.; Warner, S. J.; Lane, J. M. (1 January 2017). "7.16 Materials in Fracture Fixation☆". Comprehensive Biomaterials II. Elsevier. pp. 278–297. ISBN 978-0-08-100692-4. Retrieved 29 January 2022.
- ↑ Perren, S. M.; Russenberger, M.; Steinemann, S.; Müller, M. E.; Allgöwer, M. (1969). "A dynamic compression plate". Acta Orthopaedica Scandinavica. 40 (sup125): 29–41. doi:10.3109/ort.1969.40.suppl-125.01. PMID 5378127.
- ↑ Buckley, Richard E.; Moran, Christopher G.; Apivatthakakul, Theerachai (2017). "3.2.2 Plates". AO Principles of Fracture Management (3rd ed.). Thieme. ISBN 978-3-13-242309-1.
- ↑ "Basic principles of plating". AO Surgery Reference. AO Foundation. Retrieved 2026-04-27.
- ↑ Uhthoff, H. K.; Dubuc, F. L. (1971). "Bone structure changes in the dog under rigid internal fixation". Clinical Orthopaedics and Related Research. 81: 165–170. doi:10.1097/00003086-197111000-00026. PMID 5133036.
- ↑ Brooks, D. B.; Burstein, A. H.; Frankel, V. H. (April 1970). "The biomechanics of torsional fractures. The stress concentration effect of a drill hole". The Journal of Bone and Joint Surgery. American Volume. 52 (3): 507–514. PMID 5425646.
- ↑ Deluca, P. A.; Lindsey, R. W.; Ruwe, P. A. (October 1988). "Refracture of bones of the forearm after the removal of compression plates". The Journal of Bone and Joint Surgery. American Volume. 70 (9): 1372–1376. PMID 3182889.
- ↑ Jensen, J. S.; Sonne-Holm, S.; Tøndevold, E. (1980). "Unstable trochanteric fractures: a comparative analysis of four methods of internal fixation". Acta Orthopaedica Scandinavica. 51 (6): 949–962. doi:10.3109/17453678008990900. PMID 7211299.
- ↑ Baumgaertner, M. R.; Curtin, S. L.; Lindskog, D. M.; Keggi, J. M. (July 1995). "The value of the tip-apex distance in predicting failure of fixation of peritrochanteric fractures of the hip". The Journal of Bone and Joint Surgery. American Volume. 77 (7): 1058–1064. doi:10.2106/00004623-199507000-00012. PMID 7608228.
- ↑ Chirodian, N.; Arch, B.; Parker, M. J. (June 2005). "Sliding hip screw fixation of trochanteric hip fractures: outcome of 1024 procedures". Injury. 36 (6): 793–800. doi:10.1016/j.injury.2005.01.017. PMID 15910836.
- ↑ Brodell, J. D.; Leve, A. R. (June 1983). "Disengagement and intrapelvic protrusion of the screw from a sliding screw-plate device. A case report". The Journal of Bone and Joint Surgery. American Volume. 65 (5): 697–701. PMID 6853577.
Notes
This article is a direct transclusion of the Wikipedia article and therefore may not meet the same editing standards as LIMSwiki.