Artificial lung
Introduction
Artificial lungs are designed to replicate the functions of the native human pulmonary tissue - the oxygenation of blood and subsequent removal of carbon dioxide. They have become prominent in recent news.The only current treatment for diseased lungs is a lung transplant, which carries several risks, and is not readily available [1]. The amount of people needing a lung transplant grows each year, and the supply of available transplants is unable to keep up [2]. With this in mind, research has moved forward with artificial lungs.
A true artificial lung kept a patient alive for 48 hours until a lung transplant could be performed.[3] Artificial lungs have also been used in cases to replace animal testing, to test products such as fragrances, oils, and other aerosols.[4]
The technology has gained more popularity in news outlets as the research advances from using artificial lungs as temporary supplements to permanent replacements for organ transplants. Innovations like creating working lungs in a lab and successfully transplanting them back into a living animal [1], serve to highlight the potential applications of this technology. This technology is not a clinical product on the market - this article will focus on current research and development of the technology.
History and Development
Clinical use of oxygenators for artificial lungs have included bubble, film, and membrane type [5]. Bubble-type oxygenators directly bubble oxygen into the blood, which can cause damage to the blood [5]. Film-type oxygenators exchange gases with the environment through films, and can also lead to blood damage [5]. Membrane-type oxygenators, which can be plate or hollow fiber shaped, use a membrane to separate chambers of oxygen and blood[5].

Early attempts to create an artificial oxygenator focused on hemodialysis cellulose, cellophane, and polyethylene membranes [6]. In 1950, a bubble type oxygenator was used to support a heart-lung machine, which was successfully used in 1957[5]. Silicon based Lande-Edward and Kolobow lungs came about in 1969 and 1971, respectively[6]. Artificial lungs today typically use hollow fiber membranes to oxygenate the blood. This type of membrane entered the market in 1979 and has been associated with lower rates of blood trauma[5][6].
Additionally, extracorporeal membrane oxygenators, or ECMO, can be used temporarily to remove carbon dioxide and add oxygen to the blood stream[7][3]. ECMOs are sufficient oxygenators for up to several hours, but research into their long term use is still ongoing[6]. Issues have been seen with blood clots forming where the gas exchange is occurring[8]. ECMO differs from true artificial lungs because blood flow to the heart is not continued during their use[3]. Today, regardless of oxygenator therapy used (ECMO or artificial lung), lung transplant eventually becomes necessary after the natural lung fails[7].
Materials and Biocompatibility
Artificial lung material choice must be decided by biocompatibility of a material and its ability to prevent adsorptions that can stop gas exchange[9]. Poor biocompatibility leads to blood coagulation and the creation of a blood clot, which can lead to blockage of the artificial lung [9]. Specifically, once an oxygenator membrane comes into contact with blood, protein and fibrin may adsorb and prevent gas movement across[9]. To prevent this, varying techniques can be used to adjust the surface chemistry of oxygenators [5]. Lack of nucleophiles, low charge, and even amounts of hydrophobic and hydrophilic areas help improve biocompatibility of membranes [5].

Polymers chosen for oxygenator membranes typically have molecules of a high molecular weight, an amorphous structure, and are durable/non-brittle[5]. Polypropylene, poly-4-methylpentene-1, polydimethylsiloxane, polysulfone, polyethersulfone, polyether ether ketone, polyetherimide, polyethylene and polyvinylidene fluoride are some typical oxygenator membrane materials [5]. To enhance their biocompatibility, they are sometimes coated in silicon or heparin, which can also help prevent clotting [5]. Hollow fiber membranes sometimes face plasma leakage issues, when plasma enters the pores of the fiber and blocks oxygen exchange, which can be managed by the fabrication process used [5].
There has been research into using tissue engineering to create an artificial lung [8]. This could use stem cells from the lung to create 3D tissue models, which may be able to reduce the need for lung for lung transplants[8]. As of 2018, tissue engineered lungs were considered “far from reality”[8]. Reproducing the extracellular matrix and creating fibers that are as densely packed as a real lung have been some main challenges in this application[8]. Research has begun on decellularized pig lung, polyethylene glycol, polyvinyl alcohol, and poly glycerol sebacate as potential cell scaffolds for tissue engineered lungs[8]. Stem cells could be added to these scaffolds to begin the creation of the artificial lung[8].
| Material | Measure of Oxygen Transport | Flexural Modulus (MPa) | Glass Transition Temperature (K) | ||
| Oxygen Permeability Coefficient
(m^3(STP) m)/(s*m^2*Pa) |
Oxygen Transmission Rate
(cc/m²/day) | ||||
| Polypropylene (isotactic) | 2.12-7.73 * 10^-18[10] | N/A | 1000-1389[10] | 275.5-283.7[10] | |
| Poly-4-methylpentene-1 | 317.12 * 10^-16[10] | N/A | 1300-1800[10] | 303-323[10] | |
| Polydimethylsiloxane | 4500[10]
(Filled with 33% silica) |
N/A | Young’s modulus: .36-.87 (k) | 123.3-150[10] | |
| Polysulfone
(bisphenol A) |
1.125*10^-17 [11] | N/A | 2758[10] | 459[10] | |
| Polyethersulfone | 6*10^-18[10] | N/A | 2552[10] | 498[10] | |
| Polyether ether ketone | 6.2*10^-16 [10] | N/A | 300-3700 [10] | 410-425[10] | |
| Polyetherimide | 1*10^-19 [12] | N/A | 3300 [10] | ||
| Polyethylene: low density | N/A | 2000 - 6820 [13] | 124 - 690 [13] | ||
| Polyethylene: high density | N/A | 465 - 5580 [14] | 660 - 1500 [14] | ||
| Polyvinylidene fluoride | 2.31*10^-15 [15] | N/A | 1930-8550 [16] | 238 [10] | |
Future Directions: Bioartifical Lungs
The innovations seen in the current research are promising, but there are still challenges to overcome before artificial lungs become commonplace. These challenges include, but are not limited to “creation of ideal scaffold materials, differentiation and expansion of lung-specific cell populations and full maturation of engineered constructs to provide graft longevity after implantation in vivo” [17].
The most notable issue is intra-device thrombosis. The problem seen is that “the large blood-contracting surface area of these devices results in increased thrombogenicity relative to VADs (ventricular assist device)” [18]. Thrombogenicity is the tendency of a material that is in contact with the blood to produce a clot. This causes current clinically available systems to have a short lifetime, typically from 1 to 2 weeks before an exchange is needed[18]. The end goal of artificial lung development is considered to be “a durable, fully intracorporeal device that allows long-term support with minimal inhibition of patient mobility or ambulation” [18].
To overcome these complications, researchers are exploring regenerative medicine as well as stem cell technologies. This is not a new idea - in the early 2000s, bone marrow derived stem cells were used to demonstrate engraftment in animal models, however the results were so minimal that no significant physiological effect could be contributed to the stem cells[18]. Induced pluripotent stem cells (iPSCs) were the next considered and research is still ongoing towards this topic [19].
With the uses of iPSCs, the concept of a patient derived lung graft to reduce immune responses becomes more feasible. There has been progress in this work by the use of decellularized donor lungs, where the original cells are removed, leaving the natural extracellular matrix scaffold to be seeded with the patient’s stem cells [19]. This moves forward to the goal of providing the functional device that holds less risk of negative immune response and will provide long term support [17][18].
See also
References
- ^ a b "First Working Replacement Lung Created in Lab". Culture. 2010-06-26. Retrieved 2026-04-17.
- ^ Nolan, Heather; Wang, Dongfang; Zwischenberger, Joseph B. (2011). "Artificial lung basics: fundamental challenges, alternative designs and future innovations". Organogenesis. 7 (1): 23–27. doi:10.4161/org.7.1.14025. ISSN 1555-8592. PMC 3082030. PMID 21289479.
- ^ a b c Fieldhouse, Rachel (2026-01-29). "48 hours without lungs: artificial organ kept man alive until transplant". Nature. 650 (8101): 284. Bibcode:2026Natur.650..284F. doi:10.1038/d41586-026-00239-y. ISSN 1476-4687. PMID 41612017.
- ^ "MK firm's artificial lung brings "positive change to research'". www.bbc.com. 2024-04-02. Retrieved 2026-04-17.
- ^ a b c d e f g h i j k l Teber, Oğuz Orhun; Altinay, Ayşegül Derya; Mehrabani, Seyed Ali Naziri; Tasdemir, Reyhan Sengur; Zeytuncu, Bihter; Genceli, Esra Ateş; Dulekgurgen, Ebru; Pekkan, Kerem; Koyuncu, İsmail (2022-03-01). "Polymeric hollow fiber membrane oxygenators as artificial lungs: A review". Biochemical Engineering Journal. 180 108340. Bibcode:2022BioEJ.18008340T. doi:10.1016/j.bej.2022.108340. ISSN 1369-703X.
- ^ a b c d Ota, Kei (2010-04-01). "Advances in artificial lungs". Journal of Artificial Organs. 13 (1): 13–16. doi:10.1007/s10047-010-0492-1. ISSN 1619-0904. PMID 20177723.
- ^ a b Naito, Noritsugu; Cook, Keith; Toyoda, Yoshiya; Shigemura, Norihisa (2018-10-02). "Artificial Lungs for Lung Failure: JACC Technology Corner". Journal of the American College of Cardiology. 72 (14): 1640–1652. doi:10.1016/j.jacc.2018.07.049. ISSN 0735-1097. PMID 30261966.
- ^ a b c d e f g Petrella, Francesco; Spaggiari, Lorenzo (2018-07-27). "Artificial lung". Journal of Thoracic Disease. 10 (Suppl 20): S2329–S2332. doi:10.21037/jtd.2017.12.89. ISSN 2072-1439. PMC 6081363. PMID 30123573.
- ^ a b c Arens, Jutta; Grottke, Oliver; Haverich, Axel; Maier, Lars S.; Schmitz-Rode, Thomas; Steinseifer, Ulrich; Wendel, H. P.; Rossaint, Rolf (August 2020). "Toward a Long-Term Artificial Lung". ASAIO journal (American Society for Artificial Internal Organs: 1992). 66 (8): 847–854. doi:10.1097/MAT.0000000000001139. ISSN 1538-943X. PMC 7386861. PMID 32740342.
- ^ a b c d e f g h i j k l m n o p q r Polymer Data Handbook. Oxford University Press. 1999.
- ^ "Gas Permeability Properties of Polysulfone Membranes Containing the Mesoporous Molecular Sieve MCM-41". pubs.acs.org. doi:10.1021/cm000931+. Retrieved 2026-04-21.
- ^ "Supplier Data - Polyetherimide ( PEI ) ( Goodfellow )". AZoM. 2003-03-05. Retrieved 2026-04-21.
- ^ a b "Overview of materials for Low Density Polyethylene (LDPE), Film Grade". MatWeb. 2026-04-21. Retrieved 2026-04-21.
- ^ a b "Overview of materials of High Density Polyethylene (HDPE), Film Grade". MatWeb. 2026-04-21. Retrieved 2026-04-21.
- ^ "Kynar POLYVINYLIDENE FLUORIDE Superior Permeation Barriers" (PDF). Arkema. 2026-04-21. Retrieved 2026-04-21.
- ^ "Overview of materials for Polyvinylidendefluoride (PVDF), Molded/Extruded". MatWeb. 2026-04-21. Retrieved 2026-04-21.
- ^ a b Song, J. J.; Ott, H. C. (2012-02-01). "Bioartificial Lung Engineering". American Journal of Transplantation. 12 (2): 283–288. doi:10.1111/j.1600-6143.2011.03808.x. ISSN 1600-6135.
- ^ a b c d e Orizondo, Ryan A.; Cardounel, Arturo J.; Kormos, Robert; Sanchez, Pablo G. (2019-12-01). "Artificial Lungs: Current Status and Future Directions". Current Transplantation Reports. 6 (4): 307–315. doi:10.1007/s40472-019-00255-0. ISSN 2196-3029.
- ^ a b Weymann, Alexander; Patil, Nikhil Prakash; Sabashnikov, Anton; Korkmaz, Sevil; Li, Shiliang; Soos, Pal; Ishtok, Roland; Chaimow, Nicole; Pätzold, Ines; Czerny, Natalie; Schmack, Bastian; Popov, Aron‐Frederik; Simon, Andre Rüdiger; Karck, Matthias; Szabo, Gabor (December 2015). "Perfusion‐Decellularization of Porcine Lung and Trachea for Respiratory Bioengineering". Artificial Organs. 39 (12): 1024–1032. doi:10.1111/aor.12481. ISSN 0160-564X.
Notes
This article is a direct transclusion of the Wikipedia article and therefore may not meet the same editing standards as LIMSwiki.