! style="color:brown; background-color:#ffffee; width:700px;"| Requirement | ! style="color:brown; background-color:#ffffee; width:700px;"| Requirement | ||
|- | |- | ||
| style="padding:5px; width:350px;" |[https://www.law.cornell.edu/cfr/text/42/493.1241 42 CFR Part 493.1241]<br />[https://www.astm.org/ | | style="padding:5px; width:350px;" |[https://www.law.cornell.edu/cfr/text/42/493.1241 42 CFR Part 493.1241]<br />[https://www.astm.org/e1578-18.html ASTM E1578-18 C-1-1] | ||
| style="background-color:white;" |'''1.1''' The system should allow for sample registration prior to (e.g., preregistration) or after (e.g., registration) physical sample collection. | | style="background-color:white;" |'''1.1''' The system should allow for sample registration prior to (e.g., preregistration) or after (e.g., registration) physical sample collection. | ||
|- | |- | ||
Line 30: | Line 30: | ||
[https://www.law.cornell.edu/cfr/text/29/1910.1096 29 CFR Part 1910.1096 (e1-6)]<br /> | [https://www.law.cornell.edu/cfr/text/29/1910.1096 29 CFR Part 1910.1096 (e1-6)]<br /> | ||
[https://www.law.cornell.edu/cfr/text/42/493.1232 42 CFR Part 493.1232]<br /> | [https://www.law.cornell.edu/cfr/text/42/493.1232 42 CFR Part 493.1232]<br /> | ||
[ | [https://www.abft.org/files/ABFT_LAP_Standards_May_31_2013.pdf ABFT Accreditation Manual Sec. D]<br /> | ||
[https://www.astm.org/ | [https://www.astm.org/e1492-11r17.html ASTM E1492-11 4.1.1.6]<br /> | ||
[https://www.astm.org/ | [https://www.astm.org/e1578-18.html ASTM E1578-18 C-1-2]<br /> | ||
[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 3.2.2]<br /> | [https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 3.2.2]<br /> | ||
[https://www.epa.gov/quality/guidance-quality-assurance-project-plans-epa-qag-5 EPA QA/G-5 2.2.3]<br /> | [https://www.epa.gov/quality/guidance-quality-assurance-project-plans-epa-qag-5 EPA QA/G-5 2.2.3]<br /> | ||
[https://eur-lex.europa.eu/ | [https://eur-lex.europa.eu/eli/dir/2003/94/oj E.U. Commission Directive 2003/94/EC Article 15]<br />[https://www.oecd.org/chemicalsafety/testing/oecdseriesonprinciplesofgoodlaboratorypracticeglpandcompliancemonitoring.htm OECD GLP Principles 6.1]<br /> | ||
[https://www.ams.usda.gov/datasets/pdp/pdp-standard-operating-procedures USDA Sampling Procedures for PDP 6.4.1.2.6]<br /> | [https://www.ams.usda.gov/datasets/pdp/pdp-standard-operating-procedures USDA Sampling Procedures for PDP 6.4.1.2.6]<br /> | ||
[https://extranet.who.int/prequal/content/who-technical-report-series WHO Technical Report Series, #986, Annex 2, 17.11] | [https://extranet.who.int/prequal/content/who-technical-report-series WHO Technical Report Series, #986, Annex 2, 17.11] | ||
Line 43: | Line 43: | ||
| style="background-color:white;" |'''1.3''' The system shall provide a means to assign unique identifiers such as a National Provider Identifier (NPI) or Originating Agency Identifier (ORI) to the enacting entity and any subcontracted entity. The unique identifier should be able to appear on necessary records, documents, and reports referencing those entities. | | style="background-color:white;" |'''1.3''' The system shall provide a means to assign unique identifiers such as a National Provider Identifier (NPI) or Originating Agency Identifier (ORI) to the enacting entity and any subcontracted entity. The unique identifier should be able to appear on necessary records, documents, and reports referencing those entities. | ||
|- | |- | ||
| style="padding:5px; width:350px;" |[https://www.law.cornell.edu/cfr/text/42/493.1241 42 CFR Part 493.1241]<br />[https://www.astm.org/ | | style="padding:5px; width:350px;" |[https://www.law.cornell.edu/cfr/text/42/493.1241 42 CFR Part 493.1241]<br />[https://www.astm.org/e1578-18.html ASTM E1578-18 C-1-3] | ||
| style="background-color:white;" |'''1.4''' The system should provide a means for automatically registering samples in the system using any number of triggers such as dates, times, web requests, and intra-system methods like APIs and web services. | | style="background-color:white;" |'''1.4''' The system should provide a means for automatically registering samples in the system using any number of triggers such as dates, times, web requests, and intra-system methods like APIs and web services. | ||
|- | |- | ||
| style="padding:5px; width:350px;" |[https://www.astm.org/ | | style="padding:5px; width:350px;" |[https://www.astm.org/e1578-18.html ASTM E1578-18 C-1-4]<br />[https://www.ams.usda.gov/datasets/pdp/pdp-standard-operating-procedures USDA Sampling Procedures for PDP 6.4.1.1.3] | ||
| style="background-color:white;" |'''1.5''' The system should permit metadata like lot number, client demographics, sampling point, random selection process used, and industry-specific items to be included during registration. | | style="background-color:white;" |'''1.5''' The system should permit metadata like lot number, client demographics, sampling point, random selection process used, and industry-specific items to be included during registration. | ||
|- | |- | ||
| style="padding:5px; width:350px;" |[https://www.astm.org/ | | style="padding:5px; width:350px;" |[https://www.astm.org/e1578-18.html ASTM E1578-18 C-1-5] | ||
| style="background-color:white;" |'''1.6''' The system shall allow for the addition, modification, and removal of new, pre-defined, and custom analytical tests. | | style="background-color:white;" |'''1.6''' The system shall allow for the addition, modification, and removal of new, pre-defined, and custom analytical tests. | ||
|- | |- | ||
| style="padding:5px; width:350px;" |[https://www.astm.org/ | | style="padding:5px; width:350px;" |[https://www.astm.org/e1578-18.html ASTM E1578-18 C-1-6] | ||
| style="background-color:white;" |'''1.7''' The system should offer safety information related to a submitted sample upon registration that is relevant to the lab's location and industry. | | style="background-color:white;" |'''1.7''' The system should offer safety information related to a submitted sample upon registration that is relevant to the lab's location and industry. | ||
|- | |- | ||
| style="padding:5px; width:350px;" |[https://www.astm.org/ | | style="padding:5px; width:350px;" |[https://www.astm.org/e1578-18.html ASTM E1578-18 C-1-7] | ||
| style="background-color:white;" |'''1.8''' The system should support the creation and use of predefined metadata templates, as well as ad-hoc, single, and multiple samples. | | style="background-color:white;" |'''1.8''' The system should support the creation and use of predefined metadata templates, as well as ad-hoc, single, and multiple samples. | ||
|- | |- | ||
Line 76: | Line 76: | ||
[https://www.aafco.org/Publications/QA-QC-Guidelines-for-Feed-Laboratories AAFCO QA/QC Guidelines for Feed Laboratories Sec. 2.2–3]<br /> | [https://www.aafco.org/Publications/QA-QC-Guidelines-for-Feed-Laboratories AAFCO QA/QC Guidelines for Feed Laboratories Sec. 2.2–3]<br /> | ||
[https://www.aavld.org/accreditation-requirements-page AAVLD Requirements for an AVMDL Sec. 5.7.1.2]<br /> | [https://www.aavld.org/accreditation-requirements-page AAVLD Requirements for an AVMDL Sec. 5.7.1.2]<br /> | ||
[ | [https://www.abft.org/files/ABFT_LAP_Standards_May_31_2013.pdf ABFT Accreditation Manual Sec. D]<br /> | ||
[https://www.astm.org/ | [https://www.astm.org/e1578-18.html ASTM E1578-18 C-1-8]<br /> | ||
[https://nepis.epa.gov/Exe/ZyPDF.cgi?Dockey=30006MXP.PDF EPA 815-R-05-004 Chap. IV, Sec. 6 and 8]<br /> | [https://nepis.epa.gov/Exe/ZyPDF.cgi?Dockey=30006MXP.PDF EPA 815-R-05-004 Chap. IV, Sec. 6 and 8]<br /> | ||
[https://nepis.epa.gov/Exe/ZyPDF.cgi?Dockey=30006MXP.PDF EPA 815-R-05-004 Chap. VI, Sec. 8]<br /> | [https://nepis.epa.gov/Exe/ZyPDF.cgi?Dockey=30006MXP.PDF EPA 815-R-05-004 Chap. VI, Sec. 8]<br /> | ||
Line 84: | Line 84: | ||
[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.7.1–6]<br /> | [https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.7.1–6]<br /> | ||
[https://www.epa.gov/quality/guidance-quality-assurance-project-plans-epa-qag-5 EPA QA/G-5 2.2.2 and 2.2.3]<br /> | [https://www.epa.gov/quality/guidance-quality-assurance-project-plans-epa-qag-5 EPA QA/G-5 2.2.2 and 2.2.3]<br /> | ||
[https://eur-lex.europa.eu/ | [https://eur-lex.europa.eu/eli/dir/2003/94/oj E.U. Commission Directive 2003/94/EC Article 11.4]<br /> | ||
[https://www.iso.org/standard/56115.html ISO 15189:2012 5.7.2]<br /> | [https://www.iso.org/standard/56115.html ISO 15189:2012 5.7.2]<br /> | ||
[ | [https://www.oecd.org/chemicalsafety/testing/oecdseriesonprinciplesofgoodlaboratorypracticeglpandcompliancemonitoring.htm OECD GLP Principles 6.1]<br /> | ||
[https://www.ams.usda.gov/datasets/pdp/pdp-standard-operating-procedures USDA Sample Processing and Analysis Procedures for PDP 5.1 and 5.1.13]<br /> | [https://www.ams.usda.gov/datasets/pdp/pdp-standard-operating-procedures USDA Sample Processing and Analysis Procedures for PDP 5.1 and 5.1.13]<br /> | ||
[https://www.ams.usda.gov/datasets/pdp/pdp-standard-operating-procedures USDA Sampling Procedures for PDP 5.1]<br /> | [https://www.ams.usda.gov/datasets/pdp/pdp-standard-operating-procedures USDA Sampling Procedures for PDP 5.1]<br /> | ||
Line 92: | Line 92: | ||
| style="background-color:white;" |'''1.9''' The system shall be able to define the sample collection details for registered samples, including container size and type, number of containers, collection date and time, name of the collector, lot number, storage location, preservation method, collection methods used (standard and nonstandard), safety concerns, and retention period. | | style="background-color:white;" |'''1.9''' The system shall be able to define the sample collection details for registered samples, including container size and type, number of containers, collection date and time, name of the collector, lot number, storage location, preservation method, collection methods used (standard and nonstandard), safety concerns, and retention period. | ||
|- | |- | ||
| style="padding:5px; width:350px;" |[https://www.astm.org/ | | style="padding:5px; width:350px;" |[https://www.astm.org/e1578-18.html ASTM E1578-18 C-1-9]<br />[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 3.2.1]<br />[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.4.5] | ||
| style="background-color:white;" |'''1.10''' The system should allow for the addition of observations and descriptions to registered samples in the form of free text. | | style="background-color:white;" |'''1.10''' The system should allow for the addition of observations and descriptions to registered samples in the form of free text. | ||
|- | |- | ||
| style="padding:5px; width:350px;" |[https://www.astm.org/ | | style="padding:5px; width:350px;" |[https://www.astm.org/e1578-18.html ASTM E1578-18 C-1-10] | ||
| style="background-color:white;" |'''1.11''' The system should support the creation of user-definable default sample registration preferences and/or input screens. | | style="background-color:white;" |'''1.11''' The system should support the creation of user-definable default sample registration preferences and/or input screens. | ||
|- | |- | ||
| style="padding:5px; width:350px;" |[https://www.astm.org/ | | style="padding:5px; width:350px;" |[https://www.astm.org/e1578-18.html ASTM E1578-18 C-1-11] | ||
| style="background-color:white;" |'''1.12''' The system should allow for the recording of sample delivery details such as deliverer, location, and date and time for a preregistered sample. | | style="background-color:white;" |'''1.12''' The system should allow for the recording of sample delivery details such as deliverer, location, and date and time for a preregistered sample. | ||
|- | |- | ||
Line 108: | Line 108: | ||
[https://www.law.cornell.edu/cfr/text/42/493.1283 42 CFR Part 493.1283]<br /> | [https://www.law.cornell.edu/cfr/text/42/493.1283 42 CFR Part 493.1283]<br /> | ||
[https://www.aavld.org/accreditation-requirements-page AAVLD Requirements for an AVMDL Sec. 5.8.2]<br /> | [https://www.aavld.org/accreditation-requirements-page AAVLD Requirements for an AVMDL Sec. 5.8.2]<br /> | ||
[ | [https://www.abft.org/files/ABFT_LAP_Standards_May_31_2013.pdf ABFT Accreditation Manual D-3]<br /> | ||
[https://www.astm.org/ | [https://www.astm.org/e1578-18.html ASTM E1578-18 C-1-12]<br /> | ||
[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 3.2.2]<br /> | [https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 3.2.2]<br /> | ||
[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.1.11]<br /> | [https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.1.11]<br /> | ||
[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.4.10]<br /> | [https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.4.10]<br /> | ||
[https://www.epa.gov/quality/guidance-quality-assurance-project-plans-epa-qag-5 EPA QA/G-5 2.2.3]<br /> | [https://www.epa.gov/quality/guidance-quality-assurance-project-plans-epa-qag-5 EPA QA/G-5 2.2.3]<br /> | ||
[https://eur-lex.europa.eu/ | [https://eur-lex.europa.eu/eli/dir/2003/94/oj E.U. Commission Directive 2003/94/EC Article 15]<br /> | ||
[https://www.iso.org/standard/66912.html ISO/IEC 17025:2017 7.4.2]<br /> | [https://www.iso.org/standard/66912.html ISO/IEC 17025:2017 7.4.2]<br /> | ||
[ | [https://www.oecd.org/chemicalsafety/testing/oecdseriesonprinciplesofgoodlaboratorypracticeglpandcompliancemonitoring.htm OECD GLP Principles 6.2]<br /> | ||
[https://www.ams.usda.gov/datasets/pdp/pdp-standard-operating-procedures USDA Sample Processing and Analysis Procedures for PDP 5.1.11]<br /> | [https://www.ams.usda.gov/datasets/pdp/pdp-standard-operating-procedures USDA Sample Processing and Analysis Procedures for PDP 5.1.11]<br /> | ||
[https://www.ams.usda.gov/datasets/pdp/pdp-standard-operating-procedures USDA Sampling Procedures for PDP 6.4.1.2.5]<br /> | [https://www.ams.usda.gov/datasets/pdp/pdp-standard-operating-procedures USDA Sampling Procedures for PDP 6.4.1.2.5]<br /> | ||
Line 128: | Line 128: | ||
[https://www.law.cornell.edu/cfr/text/21/211.194 21 CFR Part 211.194]<br /> | [https://www.law.cornell.edu/cfr/text/21/211.194 21 CFR Part 211.194]<br /> | ||
[https://www.law.cornell.edu/cfr/text/42/73.17 42 CFR Part 73.17]<br /> | [https://www.law.cornell.edu/cfr/text/42/73.17 42 CFR Part 73.17]<br /> | ||
[https://www.astm.org/ | [https://www.astm.org/e1578-18.html ASTM E1578-18 C-1-13]<br /> | ||
[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.4.7]<br /> | [https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.4.7]<br /> | ||
[https://www.epa.gov/quality/guidance-quality-assurance-project-plans-epa-qag-5 EPA QA/G-5 2.2.3]<br /> | [https://www.epa.gov/quality/guidance-quality-assurance-project-plans-epa-qag-5 EPA QA/G-5 2.2.3]<br /> | ||
[ | [https://www.oecd.org/chemicalsafety/testing/oecdseriesonprinciplesofgoodlaboratorypracticeglpandcompliancemonitoring.htm OECD GLP Principles 6.1] | ||
| style="background-color:white;" |'''1.14''' The system shall provide a user-friendly means for acknowledging the physical reception of sample material in the laboratory, including date and time of reception. | | style="background-color:white;" |'''1.14''' The system shall provide a user-friendly means for acknowledging the physical reception of sample material in the laboratory, including date and time of reception. | ||
|- | |- | ||
| style="padding:5px; width:350px;" |[https://www.astm.org/ | | style="padding:5px; width:350px;" |[https://www.astm.org/e1578-18.html ASTM E1578-18 C-1-14]<br />[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.4.6] | ||
| style="background-color:white;" |'''1.15''' The system shall allow for the comparison of received samples to the sampling requirements of the customer or laboratory in order to identify variances from those requirements. | | style="background-color:white;" |'''1.15''' The system shall allow for the comparison of received samples to the sampling requirements of the customer or laboratory in order to identify variances from those requirements. | ||
|- | |- | ||
| style="padding:5px; width:350px;" | | | style="padding:5px; width:350px;" | | ||
[https://www.aavld.org/accreditation-requirements-page AAVLD Requirements for an AVMDL Sec. 5.8.3]<br /> | [https://www.aavld.org/accreditation-requirements-page AAVLD Requirements for an AVMDL Sec. 5.8.3]<br /> | ||
[https://www.astm.org/ | [https://www.astm.org/e1492-11r17.html ASTM E1492-11 4.1.1.5]<br /> | ||
[https://www.astm.org/ | [https://www.astm.org/e1492-11r17.html ASTM E1492-11 4.1.6]<br /> | ||
[https://www.astm.org/ | [https://www.astm.org/e1492-11r17.html ASTM E1492-11 4.3.3.2]<br /> | ||
[https://www.astm.org/ | [https://www.astm.org/e1578-18.html ASTM E1578-18 C-1-15]<br /> | ||
[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.4.13]<br /> | [https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.4.13]<br /> | ||
[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.11.8]<br /> | [https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.11.8]<br /> | ||
Line 149: | Line 149: | ||
| style="background-color:white;" |'''1.16''' The system should provide a means to document any undesirable or unexpected characteristics of a submitted sample. | | style="background-color:white;" |'''1.16''' The system should provide a means to document any undesirable or unexpected characteristics of a submitted sample. | ||
|- | |- | ||
| style="padding:5px; width:350px;" |[https://www.astm.org/ | | style="padding:5px; width:350px;" |[https://www.astm.org/e1578-18.html ASTM E1578-18 C-1-16]<br />[https://www.epa.gov/quality/guidance-quality-assurance-project-plans-epa-qag-5 EPA QA/G-5 2.2.2 and 2.2.3] | ||
| style="background-color:white;" |'''1.17''' The system should provide a means to document sample preparation activities for a given sample. | | style="background-color:white;" |'''1.17''' The system should provide a means to document sample preparation activities for a given sample. | ||
|- | |- | ||
Line 160: | Line 160: | ||
[https://www.law.cornell.edu/cfr/text/42/73.17 42 CFR Part 73.17]<br /> | [https://www.law.cornell.edu/cfr/text/42/73.17 42 CFR Part 73.17]<br /> | ||
[https://www.law.cornell.edu/cfr/text/42/493.1274 42 CFR Part 493.1274]<br /> | [https://www.law.cornell.edu/cfr/text/42/493.1274 42 CFR Part 493.1274]<br /> | ||
[ | [https://www.abft.org/files/ABFT_LAP_Standards_May_31_2013.pdf ABFT Accreditation Manual Sec. D-9–10]<br /> | ||
[ | [https://des.wa.gov/sites/default/files/public/documents/About/1063/RFP/Add7_Item4ASCLD.pdf ASCLD/LAB Supp. Reqs. for the Accreditation of Forensic Science Testing Laboratories 5.8.1.1]<br /> | ||
[https://www.astm.org/ | [https://www.astm.org/e1492-11r17.html ASTM E1492-11 (throughout)]<br /> | ||
[https://www.astm.org/ | [https://www.astm.org/e1578-18.html ASTM E1578-18 C-1-17]<br /> | ||
[https://nepis.epa.gov/Exe/ZyPDF.cgi?Dockey=30006MXP.PDF EPA 815-R-05-004 Chap. III, Sec. 12]<br /> | [https://nepis.epa.gov/Exe/ZyPDF.cgi?Dockey=30006MXP.PDF EPA 815-R-05-004 Chap. III, Sec. 12]<br /> | ||
[https://nepis.epa.gov/Exe/ZyPDF.cgi?Dockey=30006MXP.PDF EPA 815-R-05-004 Appendix A]<br /> | [https://nepis.epa.gov/Exe/ZyPDF.cgi?Dockey=30006MXP.PDF EPA 815-R-05-004 Appendix A]<br /> | ||
Line 187: | Line 187: | ||
! style="color:brown; background-color:#ffffee; width:700px;"| Requirement | ! style="color:brown; background-color:#ffffee; width:700px;"| Requirement | ||
|- | |- | ||
| style="padding:5px; width:500px;" |[https://www.astm.org/ | | style="padding:5px; width:500px;" |[https://www.astm.org/e1578-18.html ASTM E1578-18 C-2-1]<br />[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.11.14] | ||
| style="background-color:white;" |'''2.1''' The system shall record and maintain the chain of custody for the laboratory's standards and reagents. | | style="background-color:white;" |'''2.1''' The system shall record and maintain the chain of custody for the laboratory's standards and reagents. | ||
|- | |- | ||
| style="padding:5px; width:500px;" |[https://www.astm.org/ | | style="padding:5px; width:500px;" |[https://www.astm.org/e1578-18.html ASTM E1578-18 C-2-2]<br />[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.11.14] | ||
| style="background-color:white;" |'''2.2''' The system shall record the current and historical storage location for the laboratory's standards and reagents. | | style="background-color:white;" |'''2.2''' The system shall record the current and historical storage location for the laboratory's standards and reagents. | ||
|- | |- | ||
| style="padding:5px; width:500px;" |[https://www.astm.org/ | | style="padding:5px; width:500px;" |[https://www.astm.org/e1578-18.html ASTM E1578-18 C-2-3]<br />[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.11.14] | ||
| style="background-color:white;" |'''2.3''' The system shall allow the laboratory's previously standardized standards and reagents to be assigned a new standard value. | | style="background-color:white;" |'''2.3''' The system shall allow the laboratory's previously standardized standards and reagents to be assigned a new standard value. | ||
|- | |- | ||
| style="padding:5px; width:500px;" |[https://www.astm.org/ | | style="padding:5px; width:500px;" |[https://www.astm.org/e1578-18.html ASTM E1578-18 C-2-4]<br />[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.11.14] | ||
| style="background-color:white;" |'''2.4''' The system shall require the recording of a standard and reagent's first opening date. | | style="background-color:white;" |'''2.4''' The system shall require the recording of a standard and reagent's first opening date. | ||
|- | |- | ||
| style="padding:5px; width:500px;" |[https://www.astm.org/ | | style="padding:5px; width:500px;" |[https://www.astm.org/e1578-18.html ASTM E1578-18 C-2-5]<br />[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.11.14] | ||
| style="background-color:white;" |'''2.5''' The system shall allow only active standards and reagents to be shown as available for use. | | style="background-color:white;" |'''2.5''' The system shall allow only active standards and reagents to be shown as available for use. | ||
|- | |- | ||
| style="padding:5px; width:500px;" |[https://www.astm.org/ | | style="padding:5px; width:500px;" |[https://www.astm.org/e1578-18.html ASTM E1578-18 C-2-6]<br />[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.11.14] | ||
| style="background-color:white;" |'''2.6''' The system shall allow standards and reagents to be flagged as no longer available for use by authorized personnel. | | style="background-color:white;" |'''2.6''' The system shall allow standards and reagents to be flagged as no longer available for use by authorized personnel. | ||
|- | |- | ||
| style="padding:5px; width:500px;" |[https://www.astm.org/ | | style="padding:5px; width:500px;" |[https://www.astm.org/e1578-18.html ASTM E1578-18 C-2-7] | ||
| style="background-color:white;" |'''2.7''' The system shall allow logically associated samples to be grouped together based on associated metadata such as sample type, test method, assigned user, and status. | | style="background-color:white;" |'''2.7''' The system shall allow logically associated samples to be grouped together based on associated metadata such as sample type, test method, assigned user, and status. | ||
|- | |- | ||
| style="padding:5px; width:500px;" |[https://www.astm.org/ | | style="padding:5px; width:500px;" |[https://www.astm.org/e1578-18.html ASTM E1578-18 C-2-8]<br />[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 3.2]<br />[https://www.iso.org/standard/66912.html ISO/IEC 17025:2017 7.4.2] | ||
| style="background-color:white;" |'''2.8''' The system should allow for the accurate identification of a physical sample in the system via barcode or RFID technology. | | style="background-color:white;" |'''2.8''' The system should allow for the accurate identification of a physical sample in the system via barcode or RFID technology. | ||
|- | |- | ||
Line 216: | Line 216: | ||
[https://www.law.cornell.edu/cfr/text/21/211.84 21 CFR Part 211.84]<br /> | [https://www.law.cornell.edu/cfr/text/21/211.84 21 CFR Part 211.84]<br /> | ||
[https://www.law.cornell.edu/cfr/text/42/73.17 42 CFR Part 73.17]<br /> | [https://www.law.cornell.edu/cfr/text/42/73.17 42 CFR Part 73.17]<br /> | ||
[https://www.astm.org/ | [https://www.astm.org/e1492-11r17.html ASTM E1492-11 4.2.1]<br /> | ||
[https://www.astm.org/ | [https://www.astm.org/e1578-18.html ASTM E1578-18 C-2-9] | ||
| style="background-color:white;" |'''2.9''' The system should provide a means to manually or automatically track sample inventory amounts and status. | | style="background-color:white;" |'''2.9''' The system should provide a means to manually or automatically track sample inventory amounts and status. | ||
|- | |- | ||
| style="padding:5px; width:500px;" |[ | | style="padding:5px; width:500px;" |[https://des.wa.gov/sites/default/files/public/documents/About/1063/RFP/Add7_Item4ASCLD.pdf ASCLD/LAB Supp. Reqs. for the Accreditation of Forensic Science Testing Laboratories 5.9.1.1]<br />[https://www.astm.org/e1578-18.html ASTM E1578-18 C-2-10] | ||
| style="background-color:white;" |'''2.10''' The system shall be able to link test methods and specifications to standards and reagents. | | style="background-color:white;" |'''2.10''' The system shall be able to link test methods and specifications to standards and reagents. | ||
|- | |- | ||
Line 234: | Line 234: | ||
[https://www.law.cornell.edu/cfr/text/21/606.65 21 CFR Part 606.65 (c)]<br /> | [https://www.law.cornell.edu/cfr/text/21/606.65 21 CFR Part 606.65 (c)]<br /> | ||
[https://www.law.cornell.edu/cfr/text/21/606.151 21 CFR Part 606.151]<br /> | [https://www.law.cornell.edu/cfr/text/21/606.151 21 CFR Part 606.151]<br /> | ||
[https://www.astm.org/ | [https://www.astm.org/e1578-18.html ASTM E1578-18 C-2-11]<br /> | ||
[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 3.1.2.1]<br /> | [https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 3.1.2.1]<br /> | ||
[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.3.2]<br /> | [https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.3.2]<br /> | ||
[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.11.3.2]<br /> | [https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.11.3.2]<br /> | ||
[https://www.epa.gov/quality/guidance-quality-assurance-project-plans-epa-qag-5 EPA QA/G-5 2.2.5]<br /> | [https://www.epa.gov/quality/guidance-quality-assurance-project-plans-epa-qag-5 EPA QA/G-5 2.2.5]<br /> | ||
[https://www.wadsworth.org/sites/default/files/WebDoc/MM%20guidance%20procedure%20for%20Micro%20iDOC-120320.pdf | [https://www.wadsworth.org/sites/default/files/WebDoc/MM%20guidance%20procedure%20for%20Micro%20iDOC-120320.pdf NYDOH ELAP Medical Marijuana Microbiology Guidance]<br /> | ||
[https://www.ams.usda.gov/datasets/pdp/pdp-standard-operating-procedures USDA Administrative Procedures for the PDP 8.5]<br /> | [https://www.ams.usda.gov/datasets/pdp/pdp-standard-operating-procedures USDA Administrative Procedures for the PDP 8.5]<br /> | ||
[https:// | [https://extranet.who.int/pqweb/medicines/who-technical-report-series WHO Technical Report Series, #986, Annex 2, 17.7]<br /> | ||
[https:// | [https://extranet.who.int/pqweb/medicines/who-technical-report-series WHO Technical Report Series, #986, Annex 2, 17.22–25] | ||
| style="background-color:white;" |'''2.11''' The system shall allow samples and tests to be created and used specifically for capturing data related to unique forms of sampling and testing such as representative sampling, calibration testing, quality control testing, preventative maintenance testing, stability testing, sterility testing, compatibility testing, identity testing, proficiency testing, and service-event-related testing. | | style="background-color:white;" |'''2.11''' The system shall allow samples and tests to be created and used specifically for capturing data related to unique forms of sampling and testing such as representative sampling, calibration testing, quality control testing, preventative maintenance testing, stability testing, sterility testing, compatibility testing, identity testing, proficiency testing, and service-event-related testing. | ||
|- | |- | ||
Line 258: | Line 258: | ||
|- | |- | ||
| style="padding:5px; width:500px;" | | | style="padding:5px; width:500px;" | | ||
[https://www.law.cornell.edu/cfr/text/21/211.84 21 CFR Part 211.84 (d)]<br />[ | [https://www.law.cornell.edu/cfr/text/21/211.84 21 CFR Part 211.84 (d)]<br />[https://des.wa.gov/sites/default/files/public/documents/About/1063/RFP/Add7_Item4ASCLD.pdf ASCLD/LAB Supp. Reqs. for the Accreditation of Forensic Science Testing Laboratories 5.1.3.1]<br />[https://www.astm.org/e1578-18.html ASTM E1578-18 C-3-1]<br />[https://www.iso.org/standard/56115.html ISO 15189:2012 5.3.2.7] | ||
| style="background-color:white;" |'''3.1''' The system shall be able to record the identity of any standards and reagents used in laboratory testing, as well as for each sample or test during result entry. Linked to their identity should also be manufacturer and supplier information (such as certificates of testing); lot numbers; reception, preparation, and expiry dates; name of preparer; reliability testing results; and approval for use. | | style="background-color:white;" |'''3.1''' The system shall be able to record the identity of any standards and reagents used in laboratory testing, as well as for each sample or test during result entry. Linked to their identity should also be manufacturer and supplier information (such as certificates of testing); lot numbers; reception, preparation, and expiry dates; name of preparer; reliability testing results; and approval for use. | ||
|- | |- | ||
| style="padding:5px; width:500px;" |[https://www.law.cornell.edu/cfr/text/21/211.84 21 CFR Part 211.84 (e)]<br />[https://www.law.cornell.edu/cfr/text/42/493.1252 42 CFR Part 493.1252 (d)]<br />[https://www.astm.org/ | | style="padding:5px; width:500px;" |[https://www.law.cornell.edu/cfr/text/21/211.84 21 CFR Part 211.84 (e)]<br />[https://www.law.cornell.edu/cfr/text/42/493.1252 42 CFR Part 493.1252 (d)]<br />[https://www.astm.org/e1578-18.html ASTM E1578-18 C-3-2] | ||
| style="background-color:white;" |'''3.2''' The system shall allow users to only select approved, non-expired standards and reagents for tests and experiments. | | style="background-color:white;" |'''3.2''' The system shall allow users to only select approved, non-expired standards and reagents for tests and experiments. | ||
|- | |- | ||
| style="padding:5px; width:500px;" |[https://www.law.cornell.edu/cfr/text/42/493.1252 42 CFR Part 493.1252 (d)]<br />[https://www.astm.org/ | | style="padding:5px; width:500px;" |[https://www.law.cornell.edu/cfr/text/42/493.1252 42 CFR Part 493.1252 (d)]<br />[https://www.astm.org/e1578-18.html ASTM E1578-18 C-3-3] | ||
| style="background-color:white;" |'''3.3''' The system shall prevent standards, reagents, and other media from being used in testing if they would expire during the testing. | | style="background-color:white;" |'''3.3''' The system shall prevent standards, reagents, and other media from being used in testing if they would expire during the testing. | ||
|- | |- | ||
| style="padding:5px; width:500px;" |[https://www.astm.org/ | | style="padding:5px; width:500px;" |[https://www.astm.org/e1578-18.html ASTM E1578-18 C-3-4] | ||
| style="background-color:white;" |'''3.4''' The system shall be able to check physical, control, and specification limits for an instrument sample. | | style="background-color:white;" |'''3.4''' The system shall be able to check physical, control, and specification limits for an instrument sample. | ||
|- | |- | ||
| style="padding:5px; width:500px;" |[https://www.astm.org/ | | style="padding:5px; width:500px;" |[https://www.astm.org/e1578-18.html ASTM E1578-18 C-3-6] | ||
| style="background-color:white;" |'''3.5''' The system should allow result entry for all tests linked to a single sample, as well as for multiple samples linked to a single test. | | style="background-color:white;" |'''3.5''' The system should allow result entry for all tests linked to a single sample, as well as for multiple samples linked to a single test. | ||
|- | |- | ||
| style="padding:5px; width:500px;" |[https://www.astm.org/ | | style="padding:5px; width:500px;" |[https://www.astm.org/e1578-18.html ASTM E1578-18 C-3-7] | ||
| style="background-color:white;" |'''3.6''' The system should allow user-definable result entry methods, e.g., uploading from a spreadsheet. | | style="background-color:white;" |'''3.6''' The system should allow user-definable result entry methods, e.g., uploading from a spreadsheet. | ||
|- | |- | ||
| style="padding:5px; width:500px;" |[https://www.astm.org/ | | style="padding:5px; width:500px;" |[https://www.astm.org/e1578-18.html ASTM E1578-18 C-3-8] | ||
| style="background-color:white;" |'''3.7''' The system should allow outsourced samples to be tracked for aliquoting, chain-of-custody, results entry, and approval purposes. | | style="background-color:white;" |'''3.7''' The system should allow outsourced samples to be tracked for aliquoting, chain-of-custody, results entry, and approval purposes. | ||
|- | |- | ||
| style="padding:5px; width:500px;" |[https://www.astm.org/ | | style="padding:5px; width:500px;" |[https://www.astm.org/e1578-18.html ASTM E1578-18 C-3-9] | ||
| style="background-color:white;" |'''3.8''' The system shall allow the creation and assignment of retest workflow, while allowing the system or assigned user to determine the existence of a retesting due date and whether or not a new sample is required for the retest. | | style="background-color:white;" |'''3.8''' The system shall allow the creation and assignment of retest workflow, while allowing the system or assigned user to determine the existence of a retesting due date and whether or not a new sample is required for the retest. | ||
|- | |- | ||
| style="padding:5px; width:500px;" |[https://www.astm.org/ | | style="padding:5px; width:500px;" |[https://www.astm.org/e1578-18.html ASTM E1578-18 C-3-10] | ||
| style="background-color:white;" |'''3.9''' The system shall support, at a minimum, the floating point/real number, integer number, text, date, list, file, calculated, Boolean, and interval test result data types, allowing users to define the data type for specific test results. | | style="background-color:white;" |'''3.9''' The system shall support, at a minimum, the floating point/real number, integer number, text, date, list, file, calculated, Boolean, and interval test result data types, allowing users to define the data type for specific test results. | ||
|- | |- | ||
| style="padding:5px; width:500px;" |[https://www.aihaaccreditedlabs.org/policies AIHA-LAP Policies | | style="padding:5px; width:500px;" |[https://www.aihaaccreditedlabs.org/policies AIHA-LAP Policies 2022 2A.7.8.2]<br />[https://www.astm.org/e1578-18.html ASTM E1578-18 C-3-11] | ||
| style="background-color:white;" |'''3.10''' The system shall allow users to enter operators such as <, >, +, and - with numeric test results. | | style="background-color:white;" |'''3.10''' The system shall allow users to enter operators such as <, >, +, and - with numeric test results. | ||
|- | |- | ||
| style="padding:5px; width:500px;" |[https://www.astm.org/ | | style="padding:5px; width:500px;" |[https://www.astm.org/e1578-18.html ASTM E1578-18 C-3-12] | ||
| style="background-color:white;" |'''3.11''' The system should allow inter- and intra-assay calculations to be performed, including the use of advanced math functions. | | style="background-color:white;" |'''3.11''' The system should allow inter- and intra-assay calculations to be performed, including the use of advanced math functions. | ||
|- | |- | ||
| style="padding:5px; width:500px;" |[https://www.law.cornell.edu/cfr/text/42/493.1241 42 CFR Part 493.1241]<br />[https://www.astm.org/ | | style="padding:5px; width:500px;" |[https://www.law.cornell.edu/cfr/text/42/493.1241 42 CFR Part 493.1241]<br />[https://www.astm.org/e1578-18.html ASTM E1578-18 C-3-13]<br />[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.9.11] | ||
| style="background-color:white;" |'''3.12''' The system should effectively alert users upon entry of out-of-specification test results. | | style="background-color:white;" |'''3.12''' The system should effectively alert users upon entry of out-of-specification test results. | ||
|- | |- | ||
Line 298: | Line 298: | ||
[https://www.law.cornell.edu/cfr/text/21/212.60 21 CFR Part 212.60 (g)]<br /> | [https://www.law.cornell.edu/cfr/text/21/212.60 21 CFR Part 212.60 (g)]<br /> | ||
[https://www.law.cornell.edu/cfr/text/21/606.160 21 CFR Part 606.160]<br /> | [https://www.law.cornell.edu/cfr/text/21/606.160 21 CFR Part 606.160]<br /> | ||
[https://www.astm.org/ | [https://www.astm.org/e1578-18.html ASTM E1578-18 C-3-14]<br /> | ||
[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 3.1.2.3–4]<br /> | [https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 3.1.2.3–4]<br /> | ||
[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 3.2]<br /> | [https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 3.2]<br /> | ||
Line 306: | Line 306: | ||
[https://www.ams.usda.gov/datasets/pdp/pdp-standard-operating-procedures USDA Data and Instrumentation for PDP 5.4.2]<br /> | [https://www.ams.usda.gov/datasets/pdp/pdp-standard-operating-procedures USDA Data and Instrumentation for PDP 5.4.2]<br /> | ||
[https://www.ams.usda.gov/datasets/pdp/pdp-standard-operating-procedures USDA Data and Instrumentation for PDP 8.1.3]<br /> | [https://www.ams.usda.gov/datasets/pdp/pdp-standard-operating-procedures USDA Data and Instrumentation for PDP 8.1.3]<br /> | ||
[https:// | [https://extranet.who.int/pqweb/medicines/who-technical-report-series WHO Technical Report Series, #986, Annex 2, 15.8] | ||
| style="background-color:white;" |'''3.13''' The system shall be able to record a complete record of all data created in the course of a test or experiment, including instruments used, calculations performed, and associated graphs, charts, and spectra. The record should also be able to capture the signatures of those who performed the test or experiment, as well as those who reviewed the record for compliance purposes. | | style="background-color:white;" |'''3.13''' The system shall be able to record a complete record of all data created in the course of a test or experiment, including instruments used, calculations performed, and associated graphs, charts, and spectra. The record should also be able to capture the signatures of those who performed the test or experiment, as well as those who reviewed the record for compliance purposes. | ||
|- | |- | ||
Line 324: | Line 324: | ||
[https://www.law.cornell.edu/cfr/text/21/211.68 21 CFR Part 211.68 (b)]<br /> | [https://www.law.cornell.edu/cfr/text/21/211.68 21 CFR Part 211.68 (b)]<br /> | ||
[https://www.law.cornell.edu/cfr/text/42/493.1241 42 CFR Part 493.1241]<br /> | [https://www.law.cornell.edu/cfr/text/42/493.1241 42 CFR Part 493.1241]<br /> | ||
[https://www.astm.org/ | [https://www.astm.org/e1578-18.html ASTM E1578-18 C-4-1]<br /> | ||
[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.9.11]<br /> | [https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.9.11]<br /> | ||
[https://www.epa.gov/quality/guidance-quality-assurance-project-plans-epa-qag-5 EPA QA/G-5 2.4.1]<br /> | [https://www.epa.gov/quality/guidance-quality-assurance-project-plans-epa-qag-5 EPA QA/G-5 2.4.1]<br /> | ||
Line 331: | Line 331: | ||
|- | |- | ||
| style="padding:5px; width:500px;" | | | style="padding:5px; width:500px;" | | ||
[https://www.astm.org/ | [https://www.astm.org/e1578-18.html ASTM E1578-18 C-4-2]<br /> | ||
[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.9.2]<br /> | [https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.9.2]<br /> | ||
[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.9.11]<br /> | [https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.9.11]<br /> | ||
Line 345: | Line 345: | ||
[https://www.law.cornell.edu/cfr/text/21/211.194 21 CFR Part 211.194]<br /> | [https://www.law.cornell.edu/cfr/text/21/211.194 21 CFR Part 211.194]<br /> | ||
[https://www.law.cornell.edu/cfr/text/45/170.315 45 CFR Part 170.315 (d)]<br /> | [https://www.law.cornell.edu/cfr/text/45/170.315 45 CFR Part 170.315 (d)]<br /> | ||
[https://www.astm.org/ | [https://www.astm.org/e1578-18.html ASTM E1578-18 C-4-3]<br /> | ||
[https://www.fbi.gov/services/cjis/cjis-security-policy-resource-center CJIS Security Policy 5.4.1.1]<br /> | [https://www.fbi.gov/services/cjis/cjis-security-policy-resource-center CJIS Security Policy 5.4.1.1]<br /> | ||
[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.9.1 and 4.9.7]<br /> | [https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.9.1 and 4.9.7]<br /> | ||
[https://eur-lex.europa.eu/ | [https://eur-lex.europa.eu/eli/dir/2003/94/oj E.U. Commission Directive 2003/94/EC Article 9.2]<br /> | ||
[https:// | [https://extranet.who.int/pqweb/medicines/who-technical-report-series WHO Technical Report Series, #986, Annex 2, 15.7 and 15.9] | ||
| style="background-color:white;" |'''4.3''' The system shall accurately maintain a full audit trail for modified results. | | style="background-color:white;" |'''4.3''' The system shall accurately maintain a full audit trail for modified results. | ||
|- | |- | ||
| style="padding:5px; width:500px;" | | | style="padding:5px; width:500px;" | | ||
[https://www.law.cornell.edu/cfr/text/21/211.68 21 CFR Part 211.68 (b)]<br /> | [https://www.law.cornell.edu/cfr/text/21/211.68 21 CFR Part 211.68 (b)]<br /> | ||
[ | [https://www.abft.org/files/ABFT_LAP_Standards_May_31_2013.pdf ABFT Accreditation Manual Sec. E-28–29]<br /> | ||
[https://www.aihaaccreditedlabs.org/Policies/Pages/default.aspx AIHA-LAP Policies | [https://www.aihaaccreditedlabs.org/Policies/Pages/default.aspx AIHA-LAP Policies 2022 2A.7.7.2 and 7.7.3]<br /> | ||
[ | [https://des.wa.gov/sites/default/files/public/documents/About/1063/RFP/Add7_Item4ASCLD.pdf ASCLD/LAB Supp. Reqs. for the Accreditation of Forensic Science Testing Laboratories 4.13.2.12]<br /> | ||
[ | [https://des.wa.gov/sites/default/files/public/documents/About/1063/RFP/Add7_Item4ASCLD.pdf ASCLD/LAB Supp. Reqs. for the Accreditation of Forensic Science Testing Laboratories 5.9.4–5]<br /> | ||
[https://www.astm.org/ | [https://www.astm.org/e1578-18.html ASTM E1578-18 C-4-4]<br /> | ||
[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 3.2.6]<br /> | [https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 3.2.6]<br /> | ||
[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.9.2–3 and 4.9.11]<br /> | [https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.9.2–3 and 4.9.11]<br /> | ||
Line 371: | Line 371: | ||
|- | |- | ||
| style="padding:5px; width:500px;" | | | style="padding:5px; width:500px;" | | ||
[https://www.astm.org/ | [https://www.astm.org/e1578-18.html ASTM E1578-18 C-4-5]<br /> | ||
[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.9.11]<br /> | [https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.9.11]<br /> | ||
[https://www.epa.gov/quality/guidance-quality-assurance-project-plans-epa-qag-5 EPA QA/G-5 2.2.10]<br /> | [https://www.epa.gov/quality/guidance-quality-assurance-project-plans-epa-qag-5 EPA QA/G-5 2.2.10]<br /> | ||
Line 380: | Line 380: | ||
|- | |- | ||
| style="padding:5px; width:500px;" | | | style="padding:5px; width:500px;" | | ||
[https://www.astm.org/ | [https://www.astm.org/e1578-18.html ASTM E1578-18 C-4-6]<br /> | ||
[https://www.epa.gov/quality/guidance-quality-assurance-project-plans-epa-qag-5 EPA QA/G-5 2.2.10]<br /> | [https://www.epa.gov/quality/guidance-quality-assurance-project-plans-epa-qag-5 EPA QA/G-5 2.2.10]<br /> | ||
[https://www.epa.gov/quality/guidance-quality-assurance-project-plans-epa-qag-5 EPA QA/G-5 2.4.1]<br /> | [https://www.epa.gov/quality/guidance-quality-assurance-project-plans-epa-qag-5 EPA QA/G-5 2.4.1]<br /> | ||
Line 398: | Line 398: | ||
! style="color:brown; background-color:#ffffee; width:700px;"| Requirement | ! style="color:brown; background-color:#ffffee; width:700px;"| Requirement | ||
|- | |- | ||
| style="padding:5px; width:500px;" |[https://www.astm.org/ | | style="padding:5px; width:500px;" |[https://www.astm.org/e1578-18.html ASTM E1578-18 C-5-1]<br />[https://www.ams.usda.gov/datasets/pdp/pdp-standard-operating-procedures USDA Sample Processing and Analysis Procedures for PDP 5.9–10] | ||
| style="background-color:white;" |'''5.1''' The system shall accurately record details of a sample's final disposition. | | style="background-color:white;" |'''5.1''' The system shall accurately record details of a sample's final disposition. | ||
|- | |- | ||
| style="padding:5px; width:500px;" |[https://www.aihaaccreditedlabs.org/policies AIHA-LAP Policies | | style="padding:5px; width:500px;" |[https://www.aihaaccreditedlabs.org/policies AIHA-LAP Policies 2022 2A.7.7.2]<br />[https://www.astm.org/e1578-18.html ASTM E1578-18 C-5-2] | ||
| style="background-color:white;" |'''5.2''' The system should allow authorized personnel to view relevant metadata for results during the review and approval process, including the instruments and reagents used, raw data from instruments, and associated reports. | | style="background-color:white;" |'''5.2''' The system should allow authorized personnel to view relevant metadata for results during the review and approval process, including the instruments and reagents used, raw data from instruments, and associated reports. | ||
|- | |- | ||
| style="padding:5px; width:500px;" |[https://www.law.cornell.edu/cfr/text/21/211.160 21 CFR Part 211.160 (b)]<br />[https://www.astm.org/ | | style="padding:5px; width:500px;" |[https://www.law.cornell.edu/cfr/text/21/211.160 21 CFR Part 211.160 (b)]<br />[https://www.astm.org/e1578-18.html ASTM E1578-18 C-5-3]<br />[https://nepis.epa.gov/Exe/ZyPDF.cgi?Dockey=30006MXP.PDF EPA 815-R-05-004 Chap. IV, Sec. 6.1] | ||
| style="background-color:white;" |'''5.3''' The system should provide appropriate functions for specific sample statuses like "approved" and "rejected," including the ability to handle re-test, re-sampling, re-calculation, out-of-specification response and notification, and disposition functions. | | style="background-color:white;" |'''5.3''' The system should provide appropriate functions for specific sample statuses like "approved" and "rejected," including the ability to handle re-test, re-sampling, re-calculation, out-of-specification response and notification, and disposition functions. | ||
|- | |- | ||
Line 419: | Line 419: | ||
! style="color:brown; background-color:#ffffee; width:700px;"| Requirement | ! style="color:brown; background-color:#ffffee; width:700px;"| Requirement | ||
|- | |- | ||
| style="padding:5px; width:500px;" |[https://www.astm.org/ | | style="padding:5px; width:500px;" |[https://www.astm.org/e1578-18.html ASTM E1578-18 C-6-1] | ||
| style="background-color:white;" |'''6.1''' The system shall be able to store files in an electronic format and link them to standards, reagents, and samples tested. | | style="background-color:white;" |'''6.1''' The system shall be able to store files in an electronic format and link them to standards, reagents, and samples tested. | ||
|- | |- | ||
Line 425: | Line 425: | ||
| style="background-color:white;" |'''6.2''' The system shall be able to export data in a computer-readable (importable into a relational database) format—such as an unformatted spreadsheet, comma-separated value (CSV) file, or extensible markup language (XML) file—as well as PDF format, for reporting purposes. | | style="background-color:white;" |'''6.2''' The system shall be able to export data in a computer-readable (importable into a relational database) format—such as an unformatted spreadsheet, comma-separated value (CSV) file, or extensible markup language (XML) file—as well as PDF format, for reporting purposes. | ||
|- | |- | ||
| style="padding:5px; width:500px;" |[https://www.law.cornell.edu/cfr/text/42/493.1291 42 CFR Part 493.1291 (g) and (h)]<br />[https://www.astm.org/ | | style="padding:5px; width:500px;" |[https://www.law.cornell.edu/cfr/text/42/493.1291 42 CFR Part 493.1291 (g) and (h)]<br />[https://www.astm.org/e1578-18.html ASTM E1578-18 C-6-2]<br />[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.11.4]<br />[https://www.epa.gov/quality/guidance-quality-assurance-project-plans-epa-qag-5 EPA QA/G-5 2.4.2] | ||
| style="background-color:white;" |'''6.3''' The system shall alert users in advance to cases when the sample may be out-of-specification or may not meet expected turnaround time requirements. | | style="background-color:white;" |'''6.3''' The system shall alert users in advance to cases when the sample may be out-of-specification or may not meet expected turnaround time requirements. | ||
|- | |- | ||
| style="padding:5px; width:500px;" |[ | | style="padding:5px; width:500px;" |[https://www.abft.org/files/ABFT_LAP_Standards_May_31_2013.pdf ABFT Accreditation Manual Sec. G-2] | ||
| style="background-color:white;" |'''6.4''' The system shall allow an authorized individual to report final results in a semi-quantitative manner (e.g., "less than X mg/L") when doing so would provide relevant information to the report recipient. | | style="background-color:white;" |'''6.4''' The system shall allow an authorized individual to report final results in a semi-quantitative manner (e.g., "less than X mg/L") when doing so would provide relevant information to the report recipient. | ||
|- | |- | ||
Line 437: | Line 437: | ||
[https://www.aavld.org/accreditation-requirements-page AAVLD Requirements for an AVMDL Sec. 5.4.2.1]<br /> | [https://www.aavld.org/accreditation-requirements-page AAVLD Requirements for an AVMDL Sec. 5.4.2.1]<br /> | ||
[https://www.aavld.org/accreditation-requirements-page AAVLD Requirements for an AVMDL Sec. 5.10.2–5]<br /> | [https://www.aavld.org/accreditation-requirements-page AAVLD Requirements for an AVMDL Sec. 5.10.2–5]<br /> | ||
[https://www.aihaaccreditedlabs.org/policies AIHA-LAP Policies | [https://www.aihaaccreditedlabs.org/policies AIHA-LAP Policies 2022 2A.7.8.1]<br /> | ||
[ | [https://www.abft.org/files/ABFT_LAP_Standards_May_31_2013.pdf ABFT Accreditation Manual Sec. E-12 and -32]<br /> | ||
[https://www.astm.org/ | [https://www.astm.org/e1578-18.html ASTM E1578-18 C-6-3]<br /> | ||
[https://nepis.epa.gov/Exe/ZyPDF.cgi?Dockey=30006MXP.PDF EPA 815-R-05-004 Chap. IV, Sec. 6.6 and 8]<br /> | [https://nepis.epa.gov/Exe/ZyPDF.cgi?Dockey=30006MXP.PDF EPA 815-R-05-004 Chap. IV, Sec. 6.6 and 8]<br /> | ||
[https://nepis.epa.gov/Exe/ZyPDF.cgi?Dockey=30006MXP.PDF EPA 815-R-05-004 Chap. VI, Sec. 8]<br /> | [https://nepis.epa.gov/Exe/ZyPDF.cgi?Dockey=30006MXP.PDF EPA 815-R-05-004 Chap. VI, Sec. 8]<br /> | ||
Line 448: | Line 448: | ||
[https://www.iso.org/standard/66912.html ISO/IEC 17025:2017 7.8.2.1]<br /> | [https://www.iso.org/standard/66912.html ISO/IEC 17025:2017 7.8.2.1]<br /> | ||
[https://www.iso.org/standard/66912.html ISO/IEC 17025:2017 7.8.3.1]<br /> | [https://www.iso.org/standard/66912.html ISO/IEC 17025:2017 7.8.3.1]<br /> | ||
[https:// | [https://extranet.who.int/pqweb/medicines/who-technical-report-series WHO Technical Report Series, #986, Annex 2, 15.43] | ||
| style="background-color:white;" |'''6.5''' The system shall substantiate the status of verified results by using tools like a certificate of analysis, which shall include details like unique identifiers; analysis procedures used; reference intervals; environmental conditions; who provided the results; additional comments, opinions, and interpretations and who provided them; and applicable times and dates. | | style="background-color:white;" |'''6.5''' The system shall substantiate the status of verified results by using tools like a certificate of analysis, which shall include details like unique identifiers; analysis procedures used; reference intervals; environmental conditions; who provided the results; additional comments, opinions, and interpretations and who provided them; and applicable times and dates. | ||
|- | |- | ||
| style="padding:5px; width:500px;" |[https://www.astm.org/ | | style="padding:5px; width:500px;" |[https://www.astm.org/e1578-18.html ASTM E1578-18 C-6-4] | ||
| style="background-color:white;" |'''6.6''' The system shall allow a certificate of analysis to be generated upon the verification and approval of results and associated metadata by authorized personnel. | | style="background-color:white;" |'''6.6''' The system shall allow a certificate of analysis to be generated upon the verification and approval of results and associated metadata by authorized personnel. | ||
|- | |- | ||
| style="padding:5px; width:500px;" |[https://www.astm.org/ | | style="padding:5px; width:500px;" |[https://www.astm.org/e1578-18.html ASTM E1578-18 C-6-5]<br />[https://www.iso.org/standard/56115.html ISO 15189:2012 5.8.1] | ||
| style="background-color:white;" |'''6.7''' The system shall provide the ability to clearly and accurately generate reports for a variety of standard situations, including certificates of analysis from testing, pending samples for approval, out-of-specification samples and tests, instruments pending calibration, expired reagents to be disposed, and trend charts of laboratory performance. | | style="background-color:white;" |'''6.7''' The system shall provide the ability to clearly and accurately generate reports for a variety of standard situations, including certificates of analysis from testing, pending samples for approval, out-of-specification samples and tests, instruments pending calibration, expired reagents to be disposed, and trend charts of laboratory performance. | ||
|- | |- | ||
Line 472: | Line 472: | ||
[https://www.law.cornell.edu/cfr/text/42/73.5 42 CFR Part 73.5–6]<br /> | [https://www.law.cornell.edu/cfr/text/42/73.5 42 CFR Part 73.5–6]<br /> | ||
[https://www.law.cornell.edu/cfr/text/42/493.43 42 CFR Part 493.43 (d)]<br /> | [https://www.law.cornell.edu/cfr/text/42/493.43 42 CFR Part 493.43 (d)]<br /> | ||
[ | [https://www.abft.org/files/ABFT_LAP_Standards_May_31_2013.pdf ABFT Accreditation Manual Sec. E-7]<br /> | ||
[https://www.astm.org/ | [https://www.astm.org/e1578-18.html ASTM E1578-18 C-6-7]<br /> | ||
[https://www.fbi.gov/services/cjis/cjis-security-policy-resource-center CJIS Security Policy 5.3.1]<br /> | [https://www.fbi.gov/services/cjis/cjis-security-policy-resource-center CJIS Security Policy 5.3.1]<br /> | ||
[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.12–15]<br /> | [https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.12–15]<br /> | ||
[https://www.epa.gov/quality/guidance-quality-assurance-project-plans-epa-qag-5 EPA QA/G-5 2.3.2]<br /> | [https://www.epa.gov/quality/guidance-quality-assurance-project-plans-epa-qag-5 EPA QA/G-5 2.3.2]<br /> | ||
[https:// | [https://csrc.nist.gov/publications/detail/sp/800-53/rev-5/final NIST 800-53, Rev. 5, IR-6 and IR-6(1)]<br /> | ||
[ | [https://www.oecd.org/chemicalsafety/testing/oecdseriesonprinciplesofgoodlaboratorypracticeglpandcompliancemonitoring.htm OECD GLP Principles 9.1]<br /> | ||
[https://www.ams.usda.gov/datasets/pdp/pdp-standard-operating-procedures USDA Administrative Procedures for the PDP (throughout)] | [https://www.ams.usda.gov/datasets/pdp/pdp-standard-operating-procedures USDA Administrative Procedures for the PDP (throughout)] | ||
| style="background-color:white;" |'''6.8''' The system should, in addition to standard reports, provide the means for custom automatic and manual reporting, including but not limited to sample registration reports, work and backlog lists, laboratory performance reports, instrument reports, statistical analysis reports, regulatory reports, incident reports, chain of custody reports, quality assurance reports, service reports, inventory analysis reports, and investigator/sponsor reports. | | style="background-color:white;" |'''6.8''' The system should, in addition to standard reports, provide the means for custom automatic and manual reporting, including but not limited to sample registration reports, work and backlog lists, laboratory performance reports, instrument reports, statistical analysis reports, regulatory reports, incident reports, chain of custody reports, quality assurance reports, service reports, inventory analysis reports, and investigator/sponsor reports. | ||
Line 486: | Line 486: | ||
[https://www.aavld.org/accreditation-requirements-page AAVLD Requirements for an AVMDL Sec. 5.10.9–10]<br /> | [https://www.aavld.org/accreditation-requirements-page AAVLD Requirements for an AVMDL Sec. 5.10.9–10]<br /> | ||
[https://www.iso.org/standard/66912.html ISO/IEC 17025:2017 7.8.8]<br /> | [https://www.iso.org/standard/66912.html ISO/IEC 17025:2017 7.8.8]<br /> | ||
[ | [https://www.oecd.org/chemicalsafety/testing/oecdseriesonprinciplesofgoodlaboratorypracticeglpandcompliancemonitoring.htm OECD GLP Principles 9.1] | ||
| style="background-color:white;" |'''6.9''' The system shall clearly identify a changed, amended, or re-issued report as being such, and clearly identify any change of information and reason for change in such a report. | | style="background-color:white;" |'''6.9''' The system shall clearly identify a changed, amended, or re-issued report as being such, and clearly identify any change of information and reason for change in such a report. | ||
|- | |- | ||
|} | |} | ||
|} | |} |