| style="padding:5px; width: | | style="padding:5px; width:350px;" |[https://www.law.cornell.edu/cfr/text/42/493.1241 42 CFR Part 493.1241]<br />[https://www.astm.org/Standards/E1578.htm ASTM E1578-18 C-1-1] | ||
| style="background-color:white;" |'''1.1''' The system should allow for sample registration prior to (e.g., preregistration) or after (e.g., registration) physical sample collection. | | style="background-color:white;" |'''1.1''' The system should allow for sample registration prior to (e.g., preregistration) or after (e.g., registration) physical sample collection. | ||
|- | |- | ||
| style="padding:5px; width: | | style="padding:5px; width:350px;" | | ||
[https://www.law.cornell.edu/cfr/text/21/58.105 21 CFR Part 58.105 (c)]<br /> | [https://www.law.cornell.edu/cfr/text/21/58.105 21 CFR Part 58.105 (c)]<br /> | ||
[https://www.law.cornell.edu/cfr/text/21/211.84 21 CFR Part 211.84]<br /> | [https://www.law.cornell.edu/cfr/text/21/211.84 21 CFR Part 211.84]<br /> | ||
Line 32: | Line 32: | ||
| style="background-color:white;" |'''1.2''' The system should allow for creation and use of pre-configured (e.g., for basic clinical samples) and customizable (e.g., for regulated activities such as blood collection and storage) sample labels, with barcode support. | | style="background-color:white;" |'''1.2''' The system should allow for creation and use of pre-configured (e.g., for basic clinical samples) and customizable (e.g., for regulated activities such as blood collection and storage) sample labels, with barcode support. | ||
|- | |- | ||
| style="padding:5px; width: | | style="padding:5px; width:350px;" |[https://www.law.cornell.edu/cfr/text/45/162.410 45 CFR Part 162.410]<br />[https://www.fbi.gov/services/cjis/cjis-security-policy-resource-center CJIS Security Policy 5.6.1] | ||
| style="background-color:white;" |'''1.3''' The system shall provide a means to assign unique identifiers such as a National Provider Identifier (NPI) or Originating Agency Identifier (ORI) to the enacting entity and any subcontracted entity. The unique identifier should be able to appear on necessary records, documents, and reports referencing those entities. | | style="background-color:white;" |'''1.3''' The system shall provide a means to assign unique identifiers such as a National Provider Identifier (NPI) or Originating Agency Identifier (ORI) to the enacting entity and any subcontracted entity. The unique identifier should be able to appear on necessary records, documents, and reports referencing those entities. | ||
|- | |- | ||
| style="padding:5px; width: | | style="padding:5px; width:350px;" |[https://www.law.cornell.edu/cfr/text/42/493.1241 42 CFR Part 493.1241]<br />[https://www.astm.org/Standards/E1578.htm ASTM E1578-18 C-1-3] | ||
| style="background-color:white;" |'''1.4''' The system should provide a means for automatically registering samples in the system using any number of triggers such as dates, times, web requests, and intra-system methods like APIs and web services. | | style="background-color:white;" |'''1.4''' The system should provide a means for automatically registering samples in the system using any number of triggers such as dates, times, web requests, and intra-system methods like APIs and web services. | ||
|- | |- | ||
| style="padding:5px; width: | | style="padding:5px; width:350px;" |[https://www.astm.org/Standards/E1578.htm ASTM E1578-18 C-1-4]<br />[https://www.ams.usda.gov/datasets/pdp/pdp-standard-operating-procedures USDA Sampling Procedures for PDP 6.4.1.1.3] | ||
| style="background-color:white;" |'''1.5''' The system should permit metadata like lot number, client demographics, sampling point, random selection process used, and industry-specific items to be included during registration. | | style="background-color:white;" |'''1.5''' The system should permit metadata like lot number, client demographics, sampling point, random selection process used, and industry-specific items to be included during registration. | ||
|- | |- | ||
| style="padding:5px; width: | | style="padding:5px; width:350px;" |[https://www.astm.org/Standards/E1578.htm ASTM E1578-18 C-1-5] | ||
| style="background-color:white;" |'''1.6''' The system shall allow for the addition, modification, and removal of new, pre-defined, and custom analytical tests. | | style="background-color:white;" |'''1.6''' The system shall allow for the addition, modification, and removal of new, pre-defined, and custom analytical tests. | ||
|- | |- | ||
| style="padding:5px; width: | | style="padding:5px; width:350px;" |[https://www.astm.org/Standards/E1578.htm ASTM E1578-18 C-1-6] | ||
| style="background-color:white;" |'''1.7''' The system should offer safety information related to a submitted sample upon registration that is relevant to the lab's location and industry. | | style="background-color:white;" |'''1.7''' The system should offer safety information related to a submitted sample upon registration that is relevant to the lab's location and industry. | ||
|- | |- | ||
| style="padding:5px; width: | | style="padding:5px; width:350px;" |[https://www.astm.org/Standards/E1578.htm ASTM E1578-18 C-1-7] | ||
| style="background-color:white;" |'''1.8''' The system should support the creation and use of predefined metadata templates, as well as ad-hoc, single, and multiple samples. | | style="background-color:white;" |'''1.8''' The system should support the creation and use of predefined metadata templates, as well as ad-hoc, single, and multiple samples. | ||
|- | |- | ||
| style="padding:5px; width: | | style="padding:5px; width:350px;" | | ||
[https://www.law.cornell.edu/cfr/text/7/331.17 7 CFR Part 331.17]<br /> | [https://www.law.cornell.edu/cfr/text/7/331.17 7 CFR Part 331.17]<br /> | ||
[https://www.law.cornell.edu/cfr/text/9/121.17 9 CFR Part 121.17]<br /> | [https://www.law.cornell.edu/cfr/text/9/121.17 9 CFR Part 121.17]<br /> | ||
Line 84: | Line 84: | ||
| style="background-color:white;" |'''1.9''' The system shall be able to define the sample collection details for registered samples, including container size and type, number of containers, collection date and time, name of the collector, lot number, storage location, preservation method, collection methods used (standard and nonstandard), safety concerns, and retention period. | | style="background-color:white;" |'''1.9''' The system shall be able to define the sample collection details for registered samples, including container size and type, number of containers, collection date and time, name of the collector, lot number, storage location, preservation method, collection methods used (standard and nonstandard), safety concerns, and retention period. | ||
|- | |- | ||
| style="padding:5px; width: | | style="padding:5px; width:350px;" |[https://www.astm.org/Standards/E1578.htm ASTM E1578-18 C-1-9]<br />[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 3.2.1]<br />[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.4.5] | ||
| style="background-color:white;" |'''1.10''' The system should allow for the addition of observations and descriptions to registered samples in the form of free text. | | style="background-color:white;" |'''1.10''' The system should allow for the addition of observations and descriptions to registered samples in the form of free text. | ||
|- | |- | ||
| style="padding:5px; width: | | style="padding:5px; width:350px;" |[https://www.astm.org/Standards/E1578.htm ASTM E1578-18 C-1-10] | ||
| style="background-color:white;" |'''1.11''' The system should support the creation of user-definable default sample registration preferences and/or input screens. | | style="background-color:white;" |'''1.11''' The system should support the creation of user-definable default sample registration preferences and/or input screens. | ||
|- | |- | ||
| style="padding:5px; width: | | style="padding:5px; width:350px;" |[https://www.astm.org/Standards/E1578.htm ASTM E1578-18 C-1-11] | ||
| style="background-color:white;" |'''1.12''' The system should allow for the recording of sample delivery details such as deliverer, location, and date and time for a preregistered sample. | | style="background-color:white;" |'''1.12''' The system should allow for the recording of sample delivery details such as deliverer, location, and date and time for a preregistered sample. | ||
|- | |- | ||
| style="padding:5px; width: | | style="padding:5px; width:350px;" | | ||
[https://www.law.cornell.edu/cfr/text/21/58.107 21 CFR Part 58.107 (c)]<br /> | [https://www.law.cornell.edu/cfr/text/21/58.107 21 CFR Part 58.107 (c)]<br /> | ||
[https://www.law.cornell.edu/cfr/text/21/211.194 21 CFR Part 211.194]<br /> | [https://www.law.cornell.edu/cfr/text/21/211.194 21 CFR Part 211.194]<br /> | ||
Line 114: | Line 114: | ||
| style="background-color:white;" |'''1.13''' The system shall assign each sample registered in the system a unique identifier using methodologies such as an ID with an incrementing integer or a user-defined naming format. | | style="background-color:white;" |'''1.13''' The system shall assign each sample registered in the system a unique identifier using methodologies such as an ID with an incrementing integer or a user-defined naming format. | ||
|- | |- | ||
| style="padding:5px; width: | | style="padding:5px; width:350px;" | | ||
[https://www.law.cornell.edu/cfr/text/7/331.17 7 CFR Part 331.17]<br /> | [https://www.law.cornell.edu/cfr/text/7/331.17 7 CFR Part 331.17]<br /> | ||
[https://www.law.cornell.edu/cfr/text/9/121.17 9 CFR Part 121.17]<br /> | [https://www.law.cornell.edu/cfr/text/9/121.17 9 CFR Part 121.17]<br /> | ||
Line 126: | Line 126: | ||
| style="background-color:white;" |'''1.14''' The system shall provide a user-friendly means for acknowledging the physical reception of sample material in the laboratory, including date and time of reception. | | style="background-color:white;" |'''1.14''' The system shall provide a user-friendly means for acknowledging the physical reception of sample material in the laboratory, including date and time of reception. | ||
|- | |- | ||
| style="padding:5px; width: | | style="padding:5px; width:350px;" |[https://www.astm.org/Standards/E1578.htm ASTM E1578-18 C-1-14]<br />[https://www.epa.gov/sites/production/files/documents/erln_lab_requirements.pdf EPA ERLN Laboratory Requirements 4.4.6] | ||
| style="background-color:white;" |'''1.15''' The system shall allow for the comparison of received samples to the sampling requirements of the customer or laboratory in order to identify variances from those requirements. | | style="background-color:white;" |'''1.15''' The system shall allow for the comparison of received samples to the sampling requirements of the customer or laboratory in order to identify variances from those requirements. | ||
|- | |- | ||
| style="padding:5px; width: | | style="padding:5px; width:350px;" | | ||
[https://www.aavld.org/accreditation-requirements-page AAVLD Requirements for an AVMDL Sec. 5.8.3]<br /> | [https://www.aavld.org/accreditation-requirements-page AAVLD Requirements for an AVMDL Sec. 5.8.3]<br /> | ||
[https://www.astm.org/Standards/E1492.htm ASTM E1492-11 4.1.1.5]<br /> | [https://www.astm.org/Standards/E1492.htm ASTM E1492-11 4.1.1.5]<br /> | ||
Line 141: | Line 141: | ||
| style="background-color:white;" |'''1.16''' The system should provide a means to document any undesirable or unexpected characteristics of a submitted sample. | | style="background-color:white;" |'''1.16''' The system should provide a means to document any undesirable or unexpected characteristics of a submitted sample. | ||
|- | |- | ||
| style="padding:5px; width: | | style="padding:5px; width:350px;" |[https://www.astm.org/Standards/E1578.htm ASTM E1578-18 C-1-16]<br />[https://www.epa.gov/quality/guidance-quality-assurance-project-plans-epa-qag-5 EPA QA/G-5 2.2.2 and 2.2.3] | ||
| style="background-color:white;" |'''1.17''' The system should provide a means to document sample preparation activities for a given sample. | | style="background-color:white;" |'''1.17''' The system should provide a means to document sample preparation activities for a given sample. | ||
|- | |- | ||
| style="padding:5px; width: | | style="padding:5px; width:350px;" | | ||
[https://www.law.cornell.edu/cfr/text/7/331.11 7 CFR Part 331.11]<br /> | [https://www.law.cornell.edu/cfr/text/7/331.11 7 CFR Part 331.11]<br /> | ||
[https://www.law.cornell.edu/cfr/text/7/331.11 7 CFR Part 331.17]<br /> | [https://www.law.cornell.edu/cfr/text/7/331.11 7 CFR Part 331.17]<br /> | ||
Line 164: | Line 164: | ||
| style="background-color:white;" |'''1.18''' The system shall have the ability to maintain the chain of custody of every sample, meaning the recording of every single sample distribution step to personnel—including details such as unique identifier, name, location, date, and time—while the sample is in the laboratory’s possession. | | style="background-color:white;" |'''1.18''' The system shall have the ability to maintain the chain of custody of every sample, meaning the recording of every single sample distribution step to personnel—including details such as unique identifier, name, location, date, and time—while the sample is in the laboratory’s possession. | ||
|- | |- | ||
| style="padding:5px; width: | | style="padding:5px; width:350px;" |[https://www.law.cornell.edu/cfr/text/7/331.16 7 CFR Part 331.16]<br />[https://www.law.cornell.edu/cfr/text/9/121.16 9 CFR Part 121.16]<br />[https://www.law.cornell.edu/cfr/text/42/73.16 42 CFR Part 73.16] | ||
| style="background-color:white;" |'''1.19''' In the case of regulated samples (e.g., select agent or toxin, cannabis, etc.), the system shall also allow for the recording of transfers to other entities outside the laboratory, including details such as personnel involved, their certification numbers (if applicable), dates, times, and any other required information. | | style="background-color:white;" |'''1.19''' In the case of regulated samples (e.g., select agent or toxin, cannabis, etc.), the system shall also allow for the recording of transfers to other entities outside the laboratory, including details such as personnel involved, their certification numbers (if applicable), dates, times, and any other required information. | ||
|- | |- | ||
|} | |} | ||
|} | |} |